Features benefits and specification

Automatic Transformer Tester 5235/5236/5237 ‧5435/5436/5437
 
  ‧0.1% Basic Accuracy
‧Up to 1MHz Frequency
‧Easily Programmed, De-skills Testing, Reduces Labor Costs
‧Speedy, Increases Productivity
‧Operator Just Loads & Removes Transformers
‧Precise Accurate & Reliable Results
‧Online Statistic, RS-232
‧Secure, Tamper-proof Programs
 
Specification
Model
5235
5236
5237
Test Parameter
Transformer Scanning Turn Ratio, Phase, L, Q, Leakage L, Balance,
ACR, Cp, D, X, Y, DCR, Pin Short
Test Signals Information
Test Frequency
(0.01%, 5digits)
Turn
50Hz ~ 200kHz (10Vac Max.)
Others
20Hz~200kHz
20Hz~500kHz
20Hz~1MHz
Output Impedance 100Ω
Measurement Range
L, LK 0.00001uH~9999.99H
C 0.00001pF~999.99mF
Q, D 0.00001~99999
Z, X, R 0.00001Ω~99.9999MΩ
Y 0.01nS~99.9999S
θ -90.00o~+90.00o
DCR 0.01mΩ~99.999MΩ
Turn 0.01~99999.99turns
Pin-Short 12 pairs, between pin to pin
Basic Accuracy
L, LK, C, Z, X, Y, R, DCR 0.1% (1kHz if AC parameter)
Q, D 0.0005 (1kHz)
θ 0.03o (1kHz)
Turn 0.5% (1kHz)
Measurement Speed (Fastest)
L, LK, C, Z, X, Y, R, Q, D, θ 75ms
DCR 25ms
Turn 75ms
Judge
Transformer Scanning PASS/FAIL judge of all test parameters
output from optional HANDLER interface
Test Channel 20 Channels
Trigger Internal, Manual, External
Display 320*240 dot-matrix LCD display
Equivalent Circuit Mode Series, Parallel
Correction Function Open/Short Zeroing, Load deviation
Memory 64 instrument setups
Operation Environment Temperature :10oC ~ 40oC, Humidity :10%~90% RH
Power Consumption 140VA max.
Power Requirement 90Vac~125Vac or 190Vac~250Vac, 48Hz~62Hz
Weight Approx. 9kgs
Dimension (W*H*D) 360*150*420mm
Interface RS-232, Handler (option), USB (option)
Tel : 886-2-2698-3877.Fax : 886-2-2698-4089‧14F-6, No.79, Hsin Tai Wu Road, Sec.1, Hsi-Chih City, Taipei 221, Taiwan
E-mail: sales@microtest.com.tw‧oliverchou@microtest.com.tw