HI-POT Tester |
| Features benefits and specification |
|
|
|
| |
 |
 |
 |
 |
| 7600/7620 |
 |
 |
Key Features |
| |
8 Channels Scan (7620 Only) |
| |
Built-in PLC Remote Interface |
| |
Three Test Modes Available |
| |
RS-232 Port |
| |
Graphic LCD Display |
| |
Programming Sequence Test |
| |
Ramp Up Time |
|
|
 |
 |
 |
 |
|
|
 |
 |
 |
 |
Specification |
AC Withstand |
| Item |
Range |
Resolution |
Accuracy |
Output Voltage V
|
100~5000@10mA
|
10
|
±(3% of setting+5V) |
Output Frequency
|
50/60Hz |
|
±100ppm |
Hi-Limit Current mA
|
0.001~10 |
0.001
|
±3% |
Lo-Limit Current mA
|
0~9.999 |
0.001
|
±3% |
Dwell Time Sec
|
0.1~99.9 |
0.1
|
|
Ramp Up Time Sec
|
0.1~99.9 |
0.1 |
|
| Arc Detection |
1~9 |
|
|
DC Withstand |
| Item |
Range |
Resolution |
Accuracy |
Output Voltage V
|
100~6000@5mA
|
10
|
±(3% of setting+5V) |
Hi-Limit Current mA
|
0.001~5 |
0.001 |
±3% |
Lo-Limit Current mA
|
0~4.999 |
0.001
|
±3% |
Dwell Time Sec
|
0.1~99.9 |
0.1
|
|
Ramp Up Time Sec
|
0.1~99.9 |
0.1
|
|
| Arc Detection |
1~9 |
|
|
Insulation Resistance |
| Item |
Range |
Resolution |
Accuracy |
Output Voltage V
|
100~1000 |
10
|
±(3% of setting+5V) |
Dwell Time Sec
|
0.1~9.99 |
0.1 |
100~500V, 1~1000MΩ(±5%) |
| Hi-Limit Resistance MΩ |
1~9999 |
1 |
500~1000V, 10~5000MΩ(±5%) |
| Lo-Limit Resistance MΩ |
0~9998 |
1
|
1000V, 10~9999M(±5%) |
General |
| Item |
Range |
Memory
|
30 Sets |
PLC Remote Control
|
Test, Abort |
| PLC Single Output |
Pass, Fail, HV output |
| LCD Display |
128*64 graphic LCD |
| Verification |
AC, DC, IR |
| Dimension and Weight |
435(W)*180(H)*455(D)mm, 15KG |
| Environment |
5~40ºC, RH 0~80% |
| Input Voltage |
115/230Vac±15%, 50/60Hz±5% |
| Test channel |
8 channels(programmable) |
|
 |
 |
 |
 |
|
|
|
|
|