トランス試験機5260N

トランス試験機5260N
Features
  • Supports 20 Test Channels
  • Basic Accuracy ±0.1%
  • Open/Short Calibration Function
  • Test Frequency: 10Hz~200kHz/ 500kHz/ 1MHz
  • Meter Mode (Displays 4 Parameters)
  • 0.1 mΩ Ultra‑Low DC Resistance Measurement (5265HR | 1A)
  • Built‑in DC Withstand and Insulation Resistance Testing
  • 100 mA DC Bias Current Testing (F5620 / F5620HR)
  • USB Access for Settings, Test Data, and Screenshots
  • DCW 10–100V High‑Sensitivity Insulation Testing for Dual‑Winding Magnetic Components

The MICROTEST Transformer Tester provides comprehensive measurement of critical transformer parameters in a single instrument. Measurement items include inductance, leakage inductance, DC resistance, AC resistance, quality factor (Q), capacitance, turns ratio, and pin short‑circuit detection.It supports test frequencies of 200kHz, 500kHz, and 1MHz, with a built‑in LCR meter measurement mode. Equipped with F5620HR(P) / F5620(P) pneumatic test fixtures, the system offers 20 or 48 channels and supports 100mA DC bias testing. Customized test fixtures are available to meet specific customer requirements.

Specifications

Specification

Series  5260N
Model  5265N  5266N  5267N
Test Frequenc 10Hz~200kHz 10Hz~500kHz 10Hz~1MHz
Channel 20
Frequency Resolution 5 digits
AC Drive Level 10mV~2Vrms
Turn Ratio Frequency 50Hz~200kHz
Output Impedance 100Ω
LCR Meter Function
DC Bias Fixture 100mA (F5620)
Safety Test  DC Withstand and Insulation Resistance Testing
IR Resolution -
DC Output Voltage -
DC Leakage Current -
Measurement Parameters Inductance (Ls), Impedance (Z), Capacitance (C), Reactance (X), Conductance (G), Susceptance (B), Admittance (Y), AC Resistance, Quality Factor (Q), Dissipation Factor (D), Phase Angle (θ), Ultra‑low DCR, Leakage Inductance (Lk), Turn Ratio, Balance, Short Circuit, DC Hipot/ Leakage Current, Insulation Resistance
Parameters Measurement Ranges and Accuracy
DCR Series 5260N
Range 0.1mΩ~99.999MΩ
Accuracy 0.01Ω~ 1Ω ±(1%+0.6mΩ)@ 0.1A (CC Mode)
  0.1% (CV Mode)
Speed 20ms
L, LK

Basic Accuracy

(AC parameter : 1KHz)

0.1% Speed 20ms
C 0.1% 20ms
Q, D  0.0005 20ms
Z, X, R 0.1% 20ms
Y 0.1% 20ms
θ 0.03° 20ms
Turn-Ratio 0.5% 50ms
Pin-Short - 15ms

 

General

Measurement Mode  LCR+Transformer
Measurement Circuit  Series/ Parallel
Correction Open Circuit/ Short Circuit
Built‑in Storage Testing files 128 sets
USB Host Storage  Depend on USB capacity
Operation Auto, Manual, Remote Trigger
Interface RS-232、Handler、LAN、USB Host、EXT.I/O
PC Link Software
Power Supply Voltage:90Vac~264Vac, Fequency:47~63Hz
Power Consumption 70VA
Display  800*480 Color LCD, 7" TF
Environment Temperature:10℃~40℃, Humidity:20~90%RH
Dimension (W*H*D 358x148x343mm
Weight 4.5Kg